Open original FDA record ↗Recall record
- FDA identifier
- Z-0495-2022
- Company in FDA record
- Covidien Llc
- FDA classification date
- Jan 13, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained.
Source facts
- Action
- URGENT MEDICAL DEVICE RECALL notification letters dated 12/2021 were sent to customers. Customer Action Required: 1. Please immediately quarantine and discontinue use of the affected item code with the associated lot number listed above. 2. Please return affected product as indicated below. All products from the affected item code and associated lot number must be returned. 3. If you have distributed the product item code and lot number listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory and email it to rs.anzrecalls@medtronic.com We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please do not hesitate to contact Azi Hosseini on +61 428 464 838, or azi.hosseini@medtronic.com
- Initiated
- 2021-12-06
- Root cause
- Device Design
- Terminated
- 2023-05-11
- Distribution
- Worldwide Distribution in the countries of Australia, Austria, Belgium, Czech Republic, France, Germany, Ireland, Kazakhstan, Netherlands, Spain, United Kingdom. No US distribution.
- Recall number
- Z-0495-2022
- Recall status
- Terminated
- CDRH record ID
- 190718
- FDA classified
- 2022-01-13
- Recall event number
- 89182