Skip to content

Public FDA record

Z-0495-2022: LigaSure Blunt Tip Laparoscopic Sealer/Divider, Nano-Coated, Product Number LF1837

Recall record Z-0495-2022 for Covidien Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0495-2022
Company in FDA record
Covidien Llc
FDA classification date
Jan 13, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in treatment while an alternate device is obtained.

Source facts

Action
URGENT MEDICAL DEVICE RECALL notification letters dated 12/2021 were sent to customers. Customer Action Required: 1. Please immediately quarantine and discontinue use of the affected item code with the associated lot number listed above. 2. Please return affected product as indicated below. All products from the affected item code and associated lot number must be returned. 3. If you have distributed the product item code and lot number listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory and email it to rs.anzrecalls@medtronic.com We regret any inconvenience this may cause. We are committed to patient safety and appreciate your prompt attention to this matter. If you have any questions regarding this communication, please do not hesitate to contact Azi Hosseini on +61 428 464 838, or azi.hosseini@medtronic.com
Initiated
2021-12-06
Root cause
Device Design
Terminated
2023-05-11
Distribution
Worldwide Distribution in the countries of Australia, Austria, Belgium, Czech Republic, France, Germany, Ireland, Kazakhstan, Netherlands, Spain, United Kingdom. No US distribution.
Recall number
Z-0495-2022
Recall status
Terminated
CDRH record ID
190718
FDA classified
2022-01-13
Recall event number
89182

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗