Skip to content

Public FDA record

Z-0496-2022: Electri-Cool II Portable Cold Therapy Unit, Model Number 767, Part Number 86193

Recall record Z-0496-2022 for Gentherm Medical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0496-2022
Company in FDA record
Gentherm Medical, LLC
FDA record creation date
Jan 13, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This recall has been initiated to provide a labeling update with revised instruction how to operate the unit with Cold Therapy pads.

Source facts

Action
URGENT Medical Device Recall notification letters dated 11/19/21 were sent to customers. Pad Instructions: 1. Read and understand operation manual before connecting paCI to Etectri-Cool II. 2. Read and understand pad application instructions accompanying GENTHERM pad. 3. Connect hose to unit by inserting connectors on hose into connectors on unit. Connectors should "click" as they lock together. 4. Apply pad to patient and attach pad to hose by inserting connectors on pad into hose connectors. Normal use of the pad is to be applied in direct contact with the patient. Pad may be used over or under a patient, in the flat state or wrapped around patient. 5. Turn unit "ON" and set to temperature. 6. After pad has filled, check water level in unit. Add sterile water or water that has been passed through a filter less than or equal to 0.22 microns, only if necessary. 7. When therapy needs to be paused, disconnect the pad from the Electri-Cool II and allow the unit to continue to run. NOTE: Reconnect the pad when ready to resume therapy. 8. When therapy is no longer needed disconnect pad from hose by pressing metal clips. Dispose of pad. 9. Tum unit "OFF" at the power switch and unplug the power cord from the outlet. INSTRUCTIONS TO CUSTOMERS: 1) Immediately examine your inventory and add this letter and/or Appendix B to Electri-Cool II manual. 2) Ensure that all users are informed of the contents of this letter. If you have further distributed this product, please provide those accounts with a copy of this notice. 3) Please complete and return the enclosed response form as soon as possible to acknowledge receipt of this notification and to inform Gentherm Medical, LLC that you have performed and completed the requested actions. Return the form by e-mail to FA2021-012@gentherm.com, or mail to: Gentherm Medical, LLC 12011 Mosteller Road Cincinnati, OH 45241 This recall should be carried out to the user level. Your assistance is appreciated and necessary to reduce potent
Initiated
2021-11-22
Root cause
Device Design
Terminated
2023-05-08
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Chile, Hong Kong, Korea, UAE.
Recall number
Z-0496-2022
Recall status
Terminated
CDRH record ID
190737
FDA record created
2022-01-13
Recall event number
89193

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗