Warmer heater coil may come in contact with the filter housing, causing a small area of the plastic housing to melt
Source facts
Action
Smiths Medical ASD notified customers by letter issued Fed''X on 12/15/04. Accounts are requested not to use the device and return the unit if the the unit has not yet been placed into service. If the unit has been placed into service and operated for more than one hour then the problem will not occur. EQ-5000 unit(s) will be replaced.
Initiated
2004-12-15
Root cause
Other
Terminated
2006-06-27
Distribution
Nationwide
Foreign: Nederland BV, UK, Finland
Recall number
Z-0497-05
Recall status
Terminated
CDRH record ID
36424
FDA classified
2005-02-03
Recall event number
30703