Z-0497-2013: Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size

Recall record Z-0497-2013 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0497-2013
Company in FDA record
Medtronic Neuromodulation
FDA classification date
Dec 13, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class I

Reason stated in the source

Medtronic notified Healthcare Professionals of the impact of unapproved drugs on the performance of the SynchroMed infusion pump system. Use of unapproved drugs with SynchroMed pumps can result in an increased risk of permanent motor stall and cessation of drug infusion. UPDATE 2/1/2016: Medtronic received approval to incorporate enhancements to the SynchroMed II pump which decreases the potential for internal electrical shorting and motor corrosion, and is retrieving any unused pumps.

Source facts

Action
The firm Medtronic, sent an "Urgent medical Device Safety Notification" dated November 2012, to its customers beginning November 09, 2012. The letter described the product, problem and actions to be taken. The customers were instructed to follow the recommendations provided in the letter and complete and return the PHYSICIAN REPLY FORM via mail at Medtronic Neuromodulation Quality Assurance, 7000 Central Ave NE, Mailstop RCC150, Minneapolis, MN 55432; fax to: 800-897-3899 or email to: neuro.quality@medtronic.com. If you have questions, please contact Medtronic Neuromodulation Technical Services at 1-800-707-0933 weekdays 7am - 6pm CST. UPDATE 2/1/2016: Medtronic began notifying consignees 2/1/2016 via, mail, telephone and/or in person visit by a Medtronic Field Representative. The Medtronic Field Representatives will retrieve devices in the US and provide an Urgent Medical Device Removal letter. The letter states that device design has been updated and the Medtronic Field Representative has retrieved any devices with a Use By date on or before 5/14/2017. An account specific Customer Confirmation Form was to be completed by either the Medtronic Field Representative or the consignee. Consignees with questions can contatct their Medtronic Field Representative, or Customer Service at 1-888-854-0978.
Initiated
2012-11-09
Root cause
Nonconforming Material/Component
Terminated
2017-12-31
Distribution
Worldwide distribution: USA (nationwide) and countries including: Canada and Mexico.
Recall number
Z-0497-2013
Recall status
Terminated
CDRH record ID
114599
FDA classified
2012-12-13
Recall event number
63712

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗