Open original FDA record ↗Recall record
- FDA identifier
- Z-0497-2019
- Company in FDA record
- COVIDIEN LLC
- FDA classification date
- Dec 18, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Software Update: External USB Drive performance and its impact on Graphic User Interface (GUI) functionality and labeling of the scalar waveform displayed on GUI during ventilation
Source facts
- Action
- Medtronic notified U.S. consignees notified via Federal Express or certified mail on September 19, 2018, and the letter informs customers that Medtronic has developed a software update, Maintenance Release (MR) 5.4. Medtronic Service Engineers will load the MR5.4 software during the next scheduled or planned maintenance visit. Immediately notify all care environments in which the PB980 ventilators are used about this notification. " If your facility has distributed PB980 ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients. " Complete the attached form and return it as directed to confirm your receipt and understanding of this information. Any incidents related to this issue or if you have any questions, please contact our Technical Support Department immediately at 1-800-255-6774. Out--Side US accounts notified by Medtronic Regulatory Partners worldwide on 9/17/18 for translation in their respective country.
- Initiated
- 2018-09-18
- Root cause
- Device Design
- Terminated
- 2021-02-01
- Distribution
- Worldwide Distribution - US Nationwide Foreign: Canada Argentina Chile Japan Hong Kong Taiwan Uruguay Austria Austria Azerbaijan Belgium Bosnia And Herzegovina Bulgaria Croatia Czech Republic France Germany Greece Hungary Ireland Israel Italy Kazakhstan Kenya Kuwait Latvia Lebanon Lithuania Netherlands Oman Pakistan Poland Portugal Romania Russian Federation Saudi Arabia Slovakia Slovenia South Africa Spain Sweden Switzerland Tunisia Turkey United Arab Emirates Iran United Kingdom
- Recall number
- Z-0497-2019
- Recall status
- Terminated
- CDRH record ID
- 167618
- FDA classified
- 2018-12-18
- Recall event number
- 81167