Skip to content

Public FDA record

Z-0497-2021: ProVu Single Use Video Stylet with ET Tube, Part/Size: 038-990-080U/ 8.0, 038-990-075U/ 7.5, 038-990-070U

Recall record Z-0497-2021 for Flexicare Medical Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0497-2021
Company in FDA record
Flexicare Medical Ltd.
FDA classification date
Nov 26, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

When the video stylet cuff is over-inflated there is potential for occlusion of the corrugated section of the ETT, and increased resistance to gas flow and difficulty in ventilating the patient.

Source facts

Action
On 09/22/2020, Sales Representatives started notifying customers of this recall verbally. Customers were asked to quarantine all affected product and ensure it is not used until further notice. On 12/23/2020, Medical Device Recall notices started being mailed to customers. Customers were asked to do the following: Affected product from the lot numbers identified should be withdrawn from use and quarantined. The lot number can be found on both the outer box label and on the individual packaging. Return the Acknowledgment and Receipt Form as soon as possible. If this letter is not addressed to your facility or you have been passed the Notification from another facility, please provide the name and contact details of the facility you represent in the Acknowledgement and Receipt Form. If you are a distributor, please acknowledge receipt and understanding of this field safety notice and relay this field safety notice to all end users. On receipt of the completed Acknowledgement and Receipt Form, you will be contacted by a representative who will arrange for collection of the quarantined devices. If you have any questions, please call the firm at 949-450-9999 ext. 102, or via email at RecallUSA@flexicare.com.
Initiated
2020-09-22
Root cause
Component design/selection
Terminated
2024-07-29
Distribution
US nationwide distribution.
Recall number
Z-0497-2021
Recall status
Terminated
CDRH record ID
183816
FDA classified
2020-11-26
Recall event number
86484

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗