lmagePilot versions 1.92 and 1.93 paired with an AeroDR or Momentum Panel potential risk of incorrect length measurement/area calculation if incorrect power sequence is followed
Source facts
Action
Konica Minolta issued Urgent: Medical Device Recall lmagePilot Versions 1.92 and 1.93 Letter to Dealers on 12/8/21 and Customers on 12/15/21 via Fed'X. Letter states reason for recall, health risk and action to take:
To ensure the correct sampling pitch is obtained from the AeroDR or Momentum panel, please refer to the enclosed "lmagePilot & AeroDR Start-up Sequence and Notice before Shutting Down lmagePilot."
Konica Minolta has issued a countermeasure to minimize the risk of a potential incident. This corrective action can only be carried out on site by a Konica Minolta trained representative. Konica Minolta Healthcare or Business Partner will contact each site to schedule a visit. A Konica Minolta Representative will be contacting you within 30 days to schedule a visit. The dispatched representative will perform a check and repair process and alert the customer if any repair was needed to any of the images on the lmagePilot server. If image repair is needed, we recommend the
images that were previously sent to PACS, NAS or burned to media be exported again.
Questions contact your Konica Minolta representative or the
Konica Minolta 24-hour Technical Support Department at 1-800-945-0456.
Initiated
2021-12-03
Root cause
Under Investigation by firm
Terminated
2024-01-05
Distribution
US Nationwide distribution.
Recall number
Z-0500-2022
Recall status
Terminated
CDRH record ID
190986
FDA record created
2022-01-14
Recall event number
89275
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.