Z-0501-2022: Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300

Recall record Z-0501-2022 for Draeger Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0501-2022
Company in FDA record
Draeger Medical, Inc.
FDA record creation date
Jan 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.

Source facts

Action
On December 17, 2021, the firm notified customers via Urgent Medical Device Correction Notice. Customers were instructed to do the following: Please ensure that all users of the above-mentioned products and other concerned persons within your organization are made aware of this Urgent Medical Device Correction Notice. If you have provided the products to third parties, please forward a copy of this information. Please keep this information at least until the corrective measure has been completed. Please also complete and return the attached Customer Acknowledgment Card to confirm to us that you have received this information. A Drger Service representative will contact you to arrange a date for the correction to be performed free of charge. Until the repair has been concluded you may continue using the device under constant supervision as long as appropriate O2-monitoring is in place and a suitable alarm limit is set. Please consider a suitable buffer when adjusting the O2 flow. If you have any questions regarding the operation of your Fabius MRI Anesthesia Machine, please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437- 2437 (press 2 at the prompt, then 2, then 2 again). If you have any questions regarding this recall communication, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).
Initiated
2021-12-16
Root cause
Process control
Terminated
2023-04-17
Distribution
US Nationwide distribution in the states of NY, PA, and WI.
Recall number
Z-0501-2022
Recall status
Terminated
CDRH record ID
190921
FDA record created
2022-01-14
Recall event number
89243

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗