Z-0501-2025: OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Cata

Recall record Z-0501-2025 for OrganOx Ltd. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0501-2025
Company in FDA record
OrganOx Ltd
FDA classification date
Nov 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket

Source facts

Action
OrganOx issued Important Field Correction Notice letter on 10/15/24 via Fed'X. letter states reason for recall, health risk and action to take: As a result of the investigation into this event, OrganOx has implemented a secondary battery retention mechanism to prevent battery movement due to jarring or impact that could be related to machine transport or movement. The secondary battery retention mechanism will be installed on your machine by OrganOx service personnel no later than 15 December 2024. Recommendations: The metra in your possession continues to be safe and effective in anticipation of this correction and the machine may be used as normal. Should an unexpected rapid decrease in battery voltage occur as noted by alarm from the metra, please determine the impact to the perfusion event and take action if needed. Should this occur, please follow the OrganOx complaints reporting process and quarantine the machine until it can be serviced. Should you have questions about this important Field Correction please contact: Michael Donahue Michael.donahue@organox.com +1 732-948-2175
Initiated
2024-10-15
Root cause
Device Design
Distribution
California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.
Recall number
Z-0501-2025
Recall status
Open, Classified
CDRH record ID
210674
FDA classified
2024-11-21
Recall event number
95581

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗