Skip to content

Public FDA record

Z-0502-2010: Alphatec Spine Solanas Titanium Pedicle Screw, Part Number 63035-12

Recall record Z-0502-2010 for Alphatec Spine, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0502-2010
Company in FDA record
Alphatec Spine, Inc.
FDA classification date
Dec 14, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Alphatec Spine discovered that the affected lots were not manufactured correctly in that the screw body portion of the assembly has only 22.5 degrees of angulation in the east-west plane instead of the designed 38-40 degrees.

Source facts

Action
Alphatec Spine telephoned consignees informing them of the above mentioned Public Reason for Recall on August 5, 2009. They were asked to check their inventory to see where this product is located. If the product had not already been used Alphatec arranged to have Federal Express contact consignee for pick-up of the product(s). They were told to expect to receive the replacement(s) within 2-3 days.
Initiated
2009-08-05
Root cause
Process control
Terminated
2010-06-07
Distribution
Nationwide distribution.
Recall number
Z-0502-2010
Recall status
Terminated
CDRH record ID
84761
FDA classified
2009-12-14
Recall event number
53144

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗