The protective shield was retracted, and the blade was exposed.
Source facts
Action
The firm disseminated their safety advisory notices on 11/03/2017 by email and visit. The letter identified the affected product, problem and actions to be taken. The firm requested the following action:"Prior to opening the Tyvek tray, please visually confirm that the plastic shield is not retracted and the activation slider is fully engaged."
Initiated
2017-11-03
Root cause
Process control
Terminated
2023-05-31
Distribution
Worldwide Distribution - US Nationwide and the countries of Italy, UK, Finland, Germany, Ireland, Turkey
Recall number
Z-0502-2018
Recall status
Terminated
CDRH record ID
160525
FDA classified
2018-02-01
Recall event number
78789