Skip to content

Public FDA record

Z-0502-2024: Cooper Surgical H/S ELLIPTOSPHERE CATH 5 Fr -For administering contrast media during Hysterosalpingograph

Recall record Z-0502-2024 for CooperSurgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0502-2024
Company in FDA record
CooperSurgical, Inc.
FDA classification date
Dec 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure

Source facts

Action
CooperSurgical issued Urgent Medical Device Safety Notice dated March 22, 2023 via UPS Certified Delivery Mail. Letter states reason for recall, health risk and action to take:Our records indicate that you may have purchased the affected Product from CooperSurgical. Please inspect stock and complete the attached Acknowledgement and Receipt Form. Once completed please return the form to CooperSurgical to acknowledge receipt of the notice. If you do not have affected Product in inventory, please use the same enclosed Form to indicate that and return it to CooperSurgical so that we may document receipt of this Recall Notice. A corrective action has been initiated to ensure this failure does not reoccur. We sincerely apologize for any inconvenience caused by this notice. CooperSurgical is committed to high quality, safe and effective products. Please feel free to reach us at 203-601-5200 ext. 3300 or via email at Recall@coopersurgical.com.
Initiated
2022-03-31
Root cause
Component design/selection
Distribution
Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLAND
Recall number
Z-0502-2024
Recall status
Open, Classified
CDRH record ID
203738
FDA classified
2023-12-06
Recall event number
93239

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗