Alphatec Spine discovered that the affected lots were not manufactured correctly in that the screw body portion of the assembly has only 22.5 degrees of angulation in the east-west plane instead of the designed 38-40 degrees.
Source facts
Action
Alphatec Spine telephoned consignees informing them of the above mentioned Public Reason for Recall on August 5, 2009. They were asked to check their inventory to see where this product is located.
If the product had not already been used Alphatec arranged to have Federal Express contact consignee for pick-up of the product(s). They were told to expect to receive the replacement(s) within 2-3 days.
Initiated
2009-08-05
Root cause
Process control
Terminated
2010-06-07
Distribution
Nationwide distribution.
Recall number
Z-0503-2010
Recall status
Terminated
CDRH record ID
84770
FDA classified
2009-12-14
Recall event number
53144