Z-0503-2022: VITROS Immunodiagnostic Products High Sample Diluent A Reagent Pack

Recall record Z-0503-2022 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0503-2022
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA record creation date
Jan 18, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Lower than expected results for VITROS Immunodiagnostic Products TSH Reagent Pack when diluted with VITROS Immunodiagnostic Products High Sample Diluent A Reagent Pack Lot 2190.

Source facts

Action
Ortho clinical Diagnostics issued an Important product correction notification on November 22, 2021. Actions to be taken by the Distributors: 1. Please provide the communication (Ref. CL2021-287) to customers who were shipped VITROS HSDA Lot 2190 from your facility. 2. Do not continue to distribute VITROS HSDA Lot 2190. 3. Complete the Confirmation of Receipt form no later than November 30, 2021. 4. Please forward this notification if the product was distributed outside of your facility. If you have questions, please contact our Ortho Care" Technical Solutions Center at 1-800-421-3311. Action to be taken by the customers: 1. Discontinue using VITROS HSDA Lot 2190 if your laboratory utilizes VITROS TSH Reagent Pack. 2.Complete the Confirmation of Receipt form no later than November 30, 2021, even if no replacement product is required. If replacement is required, indicate the quantity of VITROS HSDA Lot 2190 to be replaced. 3. Please forward this notification if the product was distributed outside of your facility. Contact Information If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311
Initiated
2021-11-22
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, IL, IN, KY, LA, MD, ME, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WI, WV and the countries of Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.
Recall number
Z-0503-2022
Recall status
Open, Classified
CDRH record ID
190719
FDA record created
2022-01-18
Recall event number
89183

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗