Open original FDA record ↗Recall record
- FDA identifier
- Z-0505-2009
- Company in FDA record
- Innovative Implant Technology
- FDA classification date
- Jan 9, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Possible device failure: The cutter's shaft of the Sinu-Drill of the Sinu-Lift System could potentially get stuck inside the tap, thereby not allowing the cutter to spring out of the tap upon encountering the sinus membrane, increasing the possibility of perforating the sinus membrane.
Source facts
- Action
- All consignees will be notified by letter entitled "URGENT: MEDICAL DEVICE RECALL" dated September 30, 2008, using priority mail. Consignees will also be contacted by telephone. Consignees will be asked to return the product using the prepaid FedEx label included with the letter. The returned product will be quarantined. For further information, please telephone Innovative Implant Technology at 305-975-4101
- Initiated
- 2008-09-30
- Root cause
- Nonconforming Material/Component
- Terminated
- 2009-10-01
- Distribution
- Worldwide Distribution --- including USA and countries of Canada, Mexico, Japan, Australia, Japan, Egypt, and Turkey.
- Recall number
- Z-0505-2009
- Recall status
- Terminated
- CDRH record ID
- 74890
- FDA classified
- 2009-01-09
- Recall event number
- 50168