Failure, mechanical: If the bed mechanism is cranked downward when the bed is already in its lowest position, the springs may cause the bed to rise suddenly if the bed is empty, more slowly if occupied by a patient, or when the patient exits the bed. For further information, please contact Hill-Rom, Inc. at 800-445-3720.
Source facts
Action
Consignees were notified of the problem by letter dated 11/11/08 and instructed to position the bed at least 3 inches above the lowest bed position in order to prevent the problem, and then to only use the pendant electronic controls to raise and lower the head and foot sections of the bed, until Hill-Rom can repair the bed. The letter was entitled "Urgent Medical Device Correction."
Initiated
2008-11-11
Root cause
Other
Terminated
2009-12-31
Distribution
Nationwide.
Recall number
Z-0506-2009
Recall status
Terminated
CDRH record ID
74910
FDA classified
2009-01-16
Recall event number
50181