Open original FDA record ↗Recall record
- FDA identifier
- Z-0506-2025
- Company in FDA record
- Zyno Medical LLC
- FDA classification date
- Nov 21, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
Source facts
- Action
- Zyno Medical notified consignees on 10/18/2024 via email. Consignees were instructed that the updated notification supersedes current labeling and includes the following instructions: To maintain the health and runtime of the battery, it is recommended to plug the battery into AC Mains in for at least 6 hours every 30 days to fully charge the battery. Additionally, for scenarios where the Z800 pump is left plugged into AC Mains for extended periods of time, it is recommended to unplug for up to 5 seconds and re-plug the Z800 pump back into AC Mains at least every 30 days to activate the internal battery charger and ensure an expected operating time on battery power. Customers were also asked to provide acknowledgement of receipt.
- Initiated
- 2024-10-18
- Root cause
- Device Design
- Distribution
- This product is distributed only in the United States.
- Recall number
- Z-0506-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210742
- FDA classified
- 2024-11-21
- Recall event number
- 95601