The database may incorrectly match patient data resulting in the incorrect patient data being displayed. This could potentially present a risk to patient safety.
Source facts
Action
Microsoft notified the sole consignee by FAX on 7/22/2010 with an Urgent Medical Device Safety Information. The letter identified the affected product, the description of the problem, and actions to be taken to correct the problem, along with some additional information for the consignee. In the interim, the firm proposes to block access to patient records identified as mismatched. The firm also stated that the issue has been addressed in version 2.0. The customer should report any incidents to Michael Stokes, Director of Policy Compliance at 208-938-5397 or via e-mail at Michael.Stokes@microsoft.com.
Initiated
2010-07-27
Root cause
Software design
Terminated
2011-12-20
Distribution
Nationwide Distribution -- Wisconsin.
Recall number
Z-0507-2012
Recall status
Terminated
CDRH record ID
93895
FDA classified
2012-01-11
Recall event number
56545