Z-0507-2015: TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ (Hi:Art 5.0.1 / 5.0.2 / 5.0.3)

Recall record Z-0507-2015 for TomoTherapy Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0507-2015
Company in FDA record
TomoTherapy Incorporated
FDA classification date
Dec 12, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Accuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a jaw error occurs. This may result in an incorrect jaw position during treatment, without generating system interruption.

Source facts

Action
Accuray sent an "Urgent Field Safety Notice" dated November 21, 2014, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customers. For questions they can contact Accuray Customer Support at 1-877-668-8667 or visit the Accuray Technical Solutions Center at www.accuray.com>Services & Support or call: North America: +1.877.668.8667, Europe: +800.4141.9595, Hong Kong: +852.2247.8688, Japan: +81.3.6269.9556. Requested consignees to complete and return the Acknowledgement Form to Accuray Regulatory Affairs.
Initiated
2014-11-21
Root cause
Software design
Terminated
2015-07-17
Distribution
Worldwide Distribution - USA (nationwide) included the states of : AL, AZ, AR, CA, FL, GA, IL, IN, KS, ,Y, MD, MN, MO, NE, NM, NY, NC, OK, OR, PA, SD, TN, TX, UT, WA, WV, WI and WY., and the countries of : JAPAN, BELGIUM, NETHERLANDS, ITALY, GERMANY, KOREA, FRANCE, SWITZERLAND, ENGLAND, CANADA, MEXICO, HONG KONG, AUSTRALIA, SRI LANKA, GREECE, CHINA, PHILIPPINES, MARTINIQUE, ROMANIA, SPAIN and TURKEY.
Recall number
Z-0507-2015
Recall status
Terminated
CDRH record ID
131710
FDA classified
2014-12-12
Recall event number
69854

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗