Skip to content

Public FDA record

Z-0507-2024: Stay Safe Cap

Recall record Z-0507-2024 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0507-2024
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Dec 7, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Updating the labeling to include additional warnings and contraindications for potential side effects.

Source facts

Action
Consignees were mailed an IMPORTAND MEDICAL DEVICE INFORMATION notification dated 10/10/23 vis USPS. This notification is to bring awareness to consignees that Fresenius is updating the labeling to their Stay Safe Cap to include additional warnings and contraindications for potential side effects for P/N 05-95012. Increased exposure to povidone-iodine, which is present in the Stay Safe Cap, could lead to iodine-induced hypothyroidism and hypersensitivity. Clinicians are asked to not use these devices in known cases of povidone-iodine allergy. Fresenius is updating labeling to include these medical warnings and contraindications. Any clinical or medical questions can be directed to the firm at 1-855-616-2309. Consignees are asked to return the provided Reply Form by fax to 781-966-9796 or 781-472-7413, ATTN: Quality Department, or by email to NOTIFYRA@fmc-na.com.
Initiated
2023-10-10
Root cause
Labeling design
Distribution
US Nationwide distribution.
Recall number
Z-0507-2024
Recall status
Open, Classified
CDRH record ID
204253
FDA classified
2023-12-07
Recall event number
93396

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗