Z-0507-2025: V.A.C. Ulta Therapy Unit REF CAPULTDEV01 & ULTDEV01 The 3M" V.A.C. Ulta Negative Pressure Wound Therapy

Recall record Z-0507-2025 for KCI USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0507-2025
Company in FDA record
KCI USA, INC.
FDA classification date
Nov 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to potential system error alarm and the inability to initiate therapy as a result of broken occlusion sensors.

Source facts

Action
On 11/05/2024, the firm sent an "URGENT - MEDICAL DEVICE CORRECTION" Letter to customers informing them that due to a potentially broken occlusion sensor that the Veraflo instillation therapy units may encounter a System Error 007 alarm during boot up which may delay therapy. Customer are instructed to: 1. Locate all serial numbers and Therapy Unit Software version of V.A.C. Ulta Therapy Units. 2. Perform the steps outlined in the attached test procedure for the occlusion detection sensor 3. Record the test results on the provided Field Notice Receipt and Customer Response Form. Complete all remaining fields, sign and return this form via email to kci3mfieldactionresponse@solventum.com. Additional copies of the response form may be required based on the number of units in your facility. 4. For any therapy units with a FAIL test result, contact your Solventum representative to coordinate the repair/replacement of the occlusion detection sensor with your facility. If any additional issues are identified during the repair, you will be notified, and it will be the responsibility of the unit owner to determine whether to proceed with the additional services and associated costs. 5. Post this Medical Device Correction letter on or near the affected products to ensure continuous reminders of the field corrective action. For questions or assistance, contact Solventum representative, visit Solventum.com or call 800-275-4524.
Initiated
2024-11-05
Root cause
Use error
Distribution
Worldwide Distribution: U.S.(nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY O.U.S.(foreign) countries to: Australia, Estonia, Italy, Norway, Spain, Austria, Finland, Japan, Oman, Sweden, Bahrain, France, Jordan, Philippines, Switzerland, Brazil, Germany, Kuwait, Poland, Thailand, Canada, Greece, Malaysia, Qatar, Tunisia, Chile, Hong Kong, Malta, Romania, Turkey, China, India, Mexico, Saudi, Arabia, United Arab Emirates, Colombia, Indonesia, Morocco, Singapore, United Kingdom, Cyprus, Ireland, Netherlands, Slovenia, Czechia, Israel, New Zealand, South Africa, Vietnam, and Denmark
Recall number
Z-0507-2025
Recall status
Open, Classified
CDRH record ID
210675
FDA classified
2024-11-22
Recall event number
95580

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗