Reason stated in the source
Firm confirmed that two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. They were incorrectly laser marked with the wrong part numbers and sizes. Two lots were mixed from the following lot numbers: Part Number 64733-120 Lot Number 611884, and Part Number 64753-126 and Lot Number 611881.
Source facts
- Action
- The recall was initiated on August 16, 2007 with Alphatec Spine contacting each of the four direct distributors by telephone. This was followed with a Product Recall Notice, dated August 27, 2007, that was mailed to all affected customers. The distributors were informed that the firm was conducting a voluntary recall because two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. The firm asked the customers to return the product, and were informed that replacement product will be sent back.
- Initiated
- 2007-08-16
- Root cause
- Process control
- Terminated
- 2009-01-12
- Distribution
- Nationwide Distribution --- including states of AZ, FL, NY & TX.
- Recall number
- Z-0508-2009
- Recall status
- Terminated
- CDRH record ID
- 74679
- FDA classified
- 2008-12-31
- Recall event number
- 48177