Open original FDA record ↗Recall record
- FDA identifier
- Z-0508-2015
- Company in FDA record
- Philips Medical Systems, Inc.
- FDA classification date
- Dec 16, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In special cases, during the start-up of the current model of MobileDiagnost wDR(WmDR 1.1), an unintended exposure could occur.
Source facts
- Action
- Philips Healthcare sent an Important Electronic Product Radiation Warning letters dated April 8, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter alerts customers of the product defect and advises customers of Philips' plan to update the ATP software. A Philips Service Engineer will contact the customer when the Field Action Kit is released for installation. A customer notification letter which includes a statement that Philips Healthcare will without charge, remedy the defect or bring the product into compliance. Customers with questions about this issue can contact their local Philips representative or Philips Healthcare office, or call the Customer Care Solutions Center at 800-722-9377, Option 5. Callers should reference Field Correction Order FCO-71200109.
- Initiated
- 2014-04-08
- Root cause
- Software design
- Terminated
- 2018-05-31
- Distribution
- Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI.,and the countries of : Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iraq, Ireland, Italy, Jordan, Liberia, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Arab Emirates and United Kingdom.
- Recall number
- Z-0508-2015
- Recall status
- Terminated
- CDRH record ID
- 126535
- FDA classified
- 2014-12-16
- Recall event number
- 67869