Z-0508-2015: Intended for both adult and pediatric patients for taking diagnostic radiographic exposures of the skull,

Recall record Z-0508-2015 for Philips Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0508-2015
Company in FDA record
Philips Medical Systems, Inc.
FDA classification date
Dec 16, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In special cases, during the start-up of the current model of MobileDiagnost wDR(WmDR 1.1), an unintended exposure could occur.

Source facts

Action
Philips Healthcare sent an Important Electronic Product Radiation Warning letters dated April 8, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter alerts customers of the product defect and advises customers of Philips' plan to update the ATP software. A Philips Service Engineer will contact the customer when the Field Action Kit is released for installation. A customer notification letter which includes a statement that Philips Healthcare will without charge, remedy the defect or bring the product into compliance. Customers with questions about this issue can contact their local Philips representative or Philips Healthcare office, or call the Customer Care Solutions Center at 800-722-9377, Option 5. Callers should reference Field Correction Order FCO-71200109.
Initiated
2014-04-08
Root cause
Software design
Terminated
2018-05-31
Distribution
Worldwide Distribution - US (nationwide) Distribution including Puerto Rico and the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI.,and the countries of : Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Iraq, Ireland, Italy, Jordan, Liberia, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Portugal, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Arab Emirates and United Kingdom.
Recall number
Z-0508-2015
Recall status
Terminated
CDRH record ID
126535
FDA classified
2014-12-16
Recall event number
67869

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗