Within a specific number of Artis zee biplane and Artis Q biplane systems, a connection in the ceiling mount consisting of 12 screws may not have been attached and fastened according to quality standards. A quality inspection will be initiated to ensure all screws in the rotation unit meet quality standards.
Source facts
Action
Siemens sent an Urgent: Medical Device Correction letter dated August 24, 2017 to all affected customers. The letter identified the affected product, problem and actions to be taken. A local management or dispatch will schedule for an inspection to correct the issue. For questions please call 1-800-888-7436
Initiated
2017-08-24
Root cause
Employee error
Terminated
2018-08-29
Distribution
US Nationwide Distribution
Recall number
Z-0508-2018
Recall status
Terminated
CDRH record ID
161222
FDA record created
2018-02-03
Recall event number
79039