There is a potential for the presence of a foreign particle inside the pump motor assembly which could interfere with the motor gear rotation and lead to a permanent motor stall.
Source facts
Action
Medtronic Representatives began communicating to consignees starting 10/09/2019. Beginning 10/11/2019, Medtronic sent letters via mail. The letters explained the issue and requested the return of any unused units.
Initiated
2019-10-11
Root cause
Process control
Terminated
2023-09-15
Distribution
US and Canada, Asia -Pacific (APAC), China, Europe, Middle East and Africa (EMAE), Latin America (LATAM)
Recall number
Z-0508-2020
Recall status
Terminated
CDRH record ID
177048
FDA classified
2019-12-03
Recall event number
84037