Z-0509-2019: AIA-900 Automated Immunoassay Analyzer, product code 022930 /022930R Product Usage: The AIA-900 is capab

Recall record Z-0509-2019 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0509-2019
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Nov 28, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Tosoh Bioscience has become aware of a potential issue with the axis base of the turntable rotation drive motor on the AIA-900 Analyzer. If the tension of the timing belt between the turntable rotation pulley and drive motor is not adjusted correctly, it is possible to increase the lateral load on the motor axis resulting in the inclination and damage of the motor axis. Since January 12, 2017, Tosoh has been inspecting the tension of the timing belt during scheduled preventative maintenance visit and correcting the analyzers where this issue has been observed. If the drive motor fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received five (5) complaints related to this issue with no serious injuries reported.

Source facts

Action
On October 5, 2018, the firm issued an Urgent Medical Device Recall - Correction letter to affected customers. The letter described the product issue and advised customers that since January 12, 2017, Tosoh has been inspecting the tension of the timing belt during scheduled preventative maintenance visit and correcting the analyzers where this issue has been observed. Serial numbers of affected AIA-900 analyzers which have not yet been inspected were provided as a list. Immediate Actions to be taken by the Customer/User " Thoroughly review the content of the letter. " File this notice with your laboratory records and forward this information to other supervisors or managers in your laboratory to ensure that they are aware of the potential issue. " Complete and return the attached Acknowledgement Form by any of the listed methods. " Identify if the serial number of your AIA-900 analyzer is listed in the letter, contact Technical Support to schedule a preventative maintenance service visit. At the scheduled visit, allow the Tosoh field service engineer to inspect and verify that the axis base of the turntable rotation drive motor is operating as expected or allow the Tosoh field service engineer to correct the potential issue. " If the serial number of your analyzer is not listed, Tosoh service records indicate that the turntable rotation drive motor on your instrument(s) has already been inspected. If you have obtained unexpected test results, delays in test result reporting, or received any complaints of illness or adverse events associated with the use of the analyzer, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764 for an expedited service visit. Should you have any questions regarding this medical device recall, please feel free to contact Bernadette O'Connell at (800) 248-6764 or by email at bernadette.oconnell@tosoh.com. She will be available to answer any questions Monday - Friday, from 9:00 AM
Initiated
2018-10-05
Root cause
Nonconforming Material/Component
Terminated
2020-08-27
Distribution
Worldwide Distribution - US Nationwide in the states of FL, MD, NV, NY, SC, and TX. and Foreign distribution to Cayman Islands, Chile, Columbia, Dominican Republic, Ecuador, Guatemala, Honduras, Peru, Puerto Rico, Uruguay, and Venezuela.
Recall number
Z-0509-2019
Recall status
Terminated
CDRH record ID
167811
FDA record created
2018-11-28
Recall event number
81212

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗