Open original FDA record ↗Recall record
- FDA identifier
- Z-0509-2026
- Company in FDA record
- Medtronic Perfusion Systems
- FDA classification date
- Nov 14, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
Source facts
- Action
- Medtronic Cardiac Surgery issued an Urgent: Medical Device Correction notice to its consignees on 10/03/2025. The notice explained the issue, potential harms, and requested the consignee contact Medtronic Sales Representative to schedule the service event to their affected VitalFlow" Console. Please reference Technical Service Update (TSU) "MOD0209" when scheduling. The notice also indicated that the consignees use unaffected VitalFlow Console devices, if available, until the correction has been completed. For questions regarding the communication, please contact your Medtronic Sales Representative.
- Initiated
- 2025-10-03
- Root cause
- Software in the Use Environment
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0509-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216620
- FDA classified
- 2025-11-14
- Recall event number
- 97806