Skip to content

Public FDA record

Z-0510-2022: Voyager Rollator-intended as a mechanical walker Model Number: RLEU10xx (RLEU10BL, RLEU10PK, RLEU10WT) R

Recall record Z-0510-2022 for Compass Health Brands (Corporate Office). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0510-2022
Company in FDA record
Compass Health Brands (Corporate Office)
FDA record creation date
Jan 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Wheel spoke may crack causing the wheel to separate from the axle may result in patient injuries, including bruising and lacerations requiring stitches.

Source facts

Action
Compass Health issued Urgent Medical Device recall letter dated January 2022 on 1/11/22 issued to Distributors and Customers via email or hard copy letter. Firm posted recall to their website: https://compasshealthbrands.com/product-recall . Letter states reason for recall, health risk and action to take: Users: Users should stop using the subject products and contact Compass Health Brands. REQUIRED ACTION: 1. The Voyager Rollator casters and axles require replacement. 2. Contact Compass Health Brands at (800) 947 1728 to order replacement kits, P/N Voyager Kit. This kit contains all the parts necessary to complete replacement of the Voyager Rollator casters and axles. 3. A Voyager Caster Installment Guide will be sent as part of the Field Action Kit P/N Voyager Kit and are posted on the Compass Health Brands website. No special certification is required. 4. Complete and return the attached Consumer Field Correction Response Form to Compass Health Brands within fifteen (15) days of receipt of this field correction notification confirming your acknowledgement. Send completed form to Compass Health Brands via email at recall@compasshealthbrands.com. Complete and return this form even if you do not have affected product on hand. 5. Field correction repairs must be completed on affected units before further use of the unit. Distributors: The manufacturer has changed the wheel spoke design to increase the spoke thickness and strength. The manufacturer will supply replacement wheels with the improvement. The improved wheels will be made available to users to replace the existing wheels on the product. 1. Ensure all affected personnel are fully informed of this notice. 2. Forward this notice to your Regulatory Compliance Manager, Purchasing Manager, Customer Service Manager and Field Technicians. 3. The Voyager Rollator casters and axles require replacement. 4. Dealers/distributors or end users should contact Compass Health Brands at (800) 947-1728 and to order replacem
Initiated
2022-01-10
Root cause
Device Design
Distribution
Nationwide
Recall number
Z-0510-2022
Recall status
Open, Classified
CDRH record ID
191467
FDA record created
2022-01-19
Recall event number
89389

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗