Open original FDA record ↗Recall record
- FDA identifier
- Z-0510-2024
- Company in FDA record
- Ortho-Clinical Diagnostics, INc.
- FDA classification date
- Dec 7, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended. These affected slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased total carbon dioxide (CO2) results leading to either overdiagnosis of acidosis or missing alkalosis, both which may prompt unnecessary therapies such as IV fluids and/or bicarbonate or overdiagnosis of alkalosis which may also result in unnecessary use of IV fluids and electrolyte replacement or failure to detect acidosis and necessary treatment.
Source facts
- Action
- Ortho Clinical Diagnostics (QuidelOrtho) issued Urgent Product Correction Notification letter to Distributors and End-Users on 10.25.23. Letter states reason for recall, health risk and action to take: Discontinue using, render unusable, and discard any remaining inventory of the affected lots of VITROS ECO2 Slides from your laboratory. " If available, use an alternate, unaffected lot of VITROS ECO2 Slides for testing. " Complete and return the enclosed Confirmation of Receipt form no later than November 2, 2023. QuidelOrtho will send a replacement lot or credit your account. " Ensure your laboratory follows the calibration instructions provided in the VITROS ECO2 Slides Instructions For Use. " Please forward this notification if the affected product was distributed outside of your facility. " Save this notification with your User Documentation or post this notification by each VITROS 250/350/5,1 FS/4600/5600/XT 3400/XT 7600 System until your laboratory receives a lot of VITROS ECO2 Slides unaffected by this issue. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization (formerly Ortho Care.) We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-421-3311. Enclosure: Confirmation of Receipt Form (Ref. CL2023-248_CofR)
- Initiated
- 2023-10-25
- Root cause
- Nonconforming Material/Component
- Distribution
- Nationwide Foreign: Bermuda Canada
- Recall number
- Z-0510-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204214
- FDA classified
- 2023-12-07
- Recall event number
- 93378