Skip to content

Public FDA record

Z-0512-2012: Urine Reagent Strips (URS) 11 Parameters URS-11-100G-001 510(K) number: K970250 Devise Listing number: D

Recall record Z-0512-2012 for Teco Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0512-2012
Company in FDA record
Teco Diagnostics
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

This recall was initiated in response to findings during a FDA inspection of the firm. The firm's recall has been initiated due to improper manufacture of sub-component nitrite paper Lot 35134, Expiration Date: 01/2014. Discrepancies were found on the information recorded on formulation sheet and manufacturing log sheets. Sampling size submitted for QC testing was insufficient based upon number o

Source facts

Action
Teco Diagnostics sent an "URGENT: MEDICAL DEVICE RECALL" letter dated October 14, 2011 to affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed customers to examine their inventory for the affected product and to quarantine any product found. A Recall Return Response Form was included with the letter for customers to complete and return. Contact the firm's Technical Support Department at (800) 222-9880 for questions regarding this recall.
Initiated
2011-10-14
Root cause
Nonconforming Material/Component
Terminated
2013-04-09
Distribution
Distributed in CA and Poland.
Recall number
Z-0512-2012
Recall status
Terminated
CDRH record ID
105523
FDA classified
2012-01-11
Recall event number
60476

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗