Open original FDA record ↗Recall record
- FDA identifier
- Z-0512-2012
- Company in FDA record
- Teco Diagnostics
- FDA classification date
- Jan 11, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This recall was initiated in response to findings during a FDA inspection of the firm. The firm's recall has been initiated due to improper manufacture of sub-component nitrite paper Lot 35134, Expiration Date: 01/2014. Discrepancies were found on the information recorded on formulation sheet and manufacturing log sheets. Sampling size submitted for QC testing was insufficient based upon number o
Source facts
- Action
- Teco Diagnostics sent an "URGENT: MEDICAL DEVICE RECALL" letter dated October 14, 2011 to affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed customers to examine their inventory for the affected product and to quarantine any product found. A Recall Return Response Form was included with the letter for customers to complete and return. Contact the firm's Technical Support Department at (800) 222-9880 for questions regarding this recall.
- Initiated
- 2011-10-14
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-04-09
- Distribution
- Distributed in CA and Poland.
- Recall number
- Z-0512-2012
- Recall status
- Terminated
- CDRH record ID
- 105523
- FDA classified
- 2012-01-11
- Recall event number
- 60476