Skip to content

Public FDA record

Z-0512-2020: Rocket KCH Fetal Bladder Drainage Catheter Model: R57405

Recall record Z-0512-2020 for Rocket Medical Plc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0512-2020
Company in FDA record
Rocket Medical Plc
FDA record creation date
Nov 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A complaint was received regarding kinking of the device prior to use. If a device is kinked it cannot be used. In addition, the company became aware that the device had been used for pleural drainage, which is not included in the device indications.

Source facts

Action
Rocket Medical issued Urgent Field Safety Notice (FSN) dated July 30, 2019 advising of problem, action taken: update to IFU-Do NOT pull the tip of the catheter in an attempt to straighten it as this will induce twisting of the material and cause the catheter to kink. In addition added :This product is contraindicated for pleural drainage. Communicate this FSN to Hospital Supply Departments, Medical Device Safety Officers and those working within fetal medicine within your facility. For any devices that have been transferred to other organization(s), please ensure that details of these devices are returned to Rocket Medical and that a copy of the FSN is provided to the organization(s) to which the device has been transferred. Complete, scan and return response form, or email your responses, as soon as possible but before August 31st to cust-occ110@rocketmedical.com.
Initiated
2019-08-01
Root cause
Labeling design
Terminated
2021-11-02
Distribution
Nationwide
Recall number
Z-0512-2020
Recall status
Terminated
CDRH record ID
175798
FDA record created
2019-11-26
Recall event number
83587

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗