Skip to content

Public FDA record

Z-0512-2021: Luminesse ML PMMA, 98x20mm, REF: 5336-B1 - Product Usage: used as a milling blank in the fabrication of p

Recall record Z-0512-2021 for Talladium Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0512-2021
Company in FDA record
Talladium Inc
FDA record creation date
Dec 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Heat-cured polymethylmethacrylate (PMMA) milling blank product used to make long-term temporary crowns and bridge restorations has incorrect label. label states Part no. 5336-B1 (98x20mm). However, the item inside the box was Part no. 4302-B1 (95x20mm). An incorrect label may result in compatibility issues when setting up the disc with milling machine fixtures due to size variation.

Source facts

Action
On 10/09/20, Urgent Medical Device Recall notices were mailed and/or emailed to customers. Customers were asked to do the following: 1) Check your stock and quarantine the affected stock on hand to prevent further use. 2) Using the provided customer letter and Recall Acknowledgement Form templates, communicate this recall to any of your customers who have received product included within the scope of the recall. 3) Return product, complete the enclosed Recall Acknowledgement Form, scan and email to productrecall@talladium.com Complete the attached Acknowledgement form even if you do not have any affected stock, scan and email. The following contact information was provided: 800-221-6449 and www.TALLADIUM.com
Initiated
2020-10-09
Root cause
Labeling mix-ups
Terminated
2022-08-17
Distribution
U.S. Nationwide distribution including in the states of PA, KS, OR, FL, CA.
Recall number
Z-0512-2021
Recall status
Terminated
CDRH record ID
184197
FDA record created
2020-12-01
Recall event number
86614

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗