Skip to content

Public FDA record

Z-0513-2019: PerkinElmer QSight 210 MD Mass Spectrometer Simplicity 3Q MD v1.0 with Hot Fix 2 Software intended to ide

Recall record Z-0513-2019 for PerkinElmer Health Sciences Canada, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0513-2019
Company in FDA record
PerkinElmer Health Sciences Canada, Inc.
FDA record creation date
Nov 28, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.

Source facts

Action
Perkin Elmer issued notification on 9/24/18 customers are requested to follow instructions contained within the Urgent Medical Device Correction letter and disseminate the information to appropriate functions within their organization. Inform relevant personnel and execute one of the following two options: Qption 1 (Users who do not require Autamatic Flushinq): Deactivate the automatic flushing option in your software: 1) Navigate to the Batch Acquisition Queue tab 2) Click the Queue Options button 3) Under the "System Flush and Standby" section, DESELECT the "Enable automatic flushing" option. Option 2 (Users who require Automatic Flushinqt: Add a non data (blank or "dummy") sample to the acquisition queue to be run as your final sample (before going to Flush/standby). This will ensure that a true sample or control data are not overwritten. Contact PerkinElmer Heatth Scbnces C,arda, Bolton facility Quality Manager- SandiOrdog for further information (647. 505. 8682). The issue identified above will be addressed through a software update. PerkinElmer will contact you when the fix is available and upgrade your software.
Initiated
2018-09-24
Root cause
Software design
Terminated
2019-09-17
Distribution
FL, MN, PA India, Italy, Korea, Spain
Recall number
Z-0513-2019
Recall status
Terminated
CDRH record ID
167969
FDA record created
2018-11-28
Recall event number
81266

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗