Under certain conditions when using PerkinElmer QSight 210 MD mass spectrometer with Simplicity 3Q MD and Hotfix 2 software, it is possible for the original data from the last acquired sample to be over-written when the automatic flush is enabled.
Source facts
Action
Perkin Elmer issued notification on 9/24/18 customers are requested to follow instructions contained within the Urgent Medical Device Correction letter and disseminate the information to appropriate functions within their organization.
Inform relevant personnel and execute one of the following two options:
Qption 1 (Users who do not require Autamatic Flushinq):
Deactivate the automatic flushing option in your software:
1) Navigate to the Batch Acquisition Queue tab
2) Click the Queue Options button
3) Under the "System Flush and Standby" section, DESELECT the "Enable automatic flushing" option.
Option 2 (Users who require Automatic Flushinqt:
Add a non data (blank or "dummy") sample to the acquisition queue to be run as your final sample
(before going to Flush/standby). This will ensure that a true sample or control data are not overwritten.
Contact PerkinElmer Heatth Scbnces C,arda, Bolton facility Quality Manager- SandiOrdog for further information (647. 505. 8682). The issue identified above will be addressed through a software update. PerkinElmer will contact
you when the fix is available and upgrade your software.
Initiated
2018-09-24
Root cause
Software design
Terminated
2019-09-17
Distribution
FL, MN, PA
India, Italy, Korea, Spain
Recall number
Z-0513-2019
Recall status
Terminated
CDRH record ID
167969
FDA record created
2018-11-28
Recall event number
81266