Class II

Reason stated in the source

Incomplete seal and premature expiration of individually packaged Blood Glucose Test Strips.

Source facts

Action
Medline Industries, Inc. notified customers on about 09/18/2019, via "URGENT RECALL IMMEDIATE ACTION REQUIRED" letter. Customers were informed that the heat activated adhesive used to seal the foil pouch was not fully cured during the manufacturing process. As a result, an incomplete seal and premature expiration of the individually packaged Blood Glucose Test Strips has occurred. Inspection using EVENCARE G3 control solution and unused Blood Glucose Test Strips, from the affected lot, has shown that the test strips can produce inaccurate blood glucose readings. Instructions included to immediately check inventory for the affected item and lot number, quarantine affected product, complete and return the enclosed response form, arrange for the return of all affected product, and notify customers if the product was further distributed. Questions can be directed to 866-359-1704.
Initiated
2019-09-18
Root cause
Packaging process control
Terminated
2022-05-06
Distribution
Nationwide distribution to CA, CT, GA, ID, IL, IN, KS, MA, MI, MN, MO, MS, NV, OH, OR, SC, TN, TX, UT, WA, WI, WY.
Recall number
Z-0513-2020
Recall status
Terminated
CDRH record ID
176684
FDA record created
2019-11-22
Recall event number
83890

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗