Z-0513-2021: The Biomeme SARS-CoV-2 Real-Time RT-PCR Test is a molecular in vitro diagnostic test that is based on wid

Recall record Z-0513-2021 for Biomeme, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0513-2021
Company in FDA record
Biomeme, Inc.
FDA record creation date
Dec 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware of nine reports by a single customer that the software made false positive calls. The investigation determined that the root cause is the higher than expected amber-to-red co-excitation in the assay.

Source facts

Action
On 10/23/2020, the firm pushed the protocol update and the customer who initiated the complaint was notified. All other customers were informed of the update via email on 10/27/2020. Customers were told that the firm had deployed a change to their data processing algorithm specific to the SARS-CoV-2 protocol, which comprised of increasing the value of the co-excitation filter. Instruction to Customers: If your phone is not connected to the internet via wifi or cellular network, please connect the phone and login to the Biomeme Go app. The update will sync automatically to the Biomeme Go app and requires no further action if the phone running the Biomeme Go app is connected to wifi or a cellular network. The Before Fix Revision is 2020-03-24. The After Fix Revision is 2020-10-23. The firm also emphasized the importance to manually reviewing all results in the app or on the Biomeme Cloud. Should your PCR curves look at all suspect, contact the firm at support@biomeme.com. All positive samples should be rerun.
Initiated
2020-10-23
Root cause
Software design
Terminated
2021-01-25
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, IN, NY, OH, PA, and the countries of Canada, Singapore, Zimbabwe, and Dominican Republic.
Recall number
Z-0513-2021
Recall status
Terminated
CDRH record ID
184462
FDA record created
2020-12-01
Recall event number
86757

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗