Skip to content

Public FDA record

Z-0513-2024: Sterile Water/Saline: Brand Name: Product Name/Product Code: CARDINAL: 100ML STERILE SALINE BOTTLE/1020,

Recall record Z-0513-2024 for Nurse Assist, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0513-2024
Company in FDA record
Nurse Assist, LLC
FDA classification date
Dec 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Sodium chloride irrigation USP, and sterile water for irrigation USP, that may be packaged inside kits, have the potential for a lack of sterility assurance, which could result in a nonsterile product, use of which could cause infection.

Source facts

Action
On 11/6/23, the firm issued a press release. On 11/09/23, "URGENT: MEDICAL DEVICE RECALL" notices and response forms were emailed and mailed to direct customers and distributors, who were provided with a separate customer recall notice they could forward to their customers. Customers were asked to do the following: 1) Quarantine and then return affected devices. 2) If affected products were further distributed, send these customers the attached recall notice that asks customers to return affected devices to their distributors. 3) Distributors are asked to follow up with customers to ensure they received the recall notification and are taking appropriate action. 4) Distributors are asked to provide their Customer Shipment Report that lists all devices shipped to customers. The report should include the Shipment Date, Customer Name and Address, Part Number, Lot Number, and Quantity. This information is important to determine removal effectiveness. 5) Complete and return the response form via email to nurseassist@rqa-inc.com 6) If your company repacked these medical devices into convenience kits, please contact your local FDA recall coordinator to evaluate the need for a new recall. The firm is arranging for removal and replacement of the affected product. The recall notice sent by distributors to their customers also asks that the notice be forwarded to the applicable personnel in their organization or to any organization where the Affected Devices were transferred. Customers with questions can call 800-649-6800, between the hours of 8am and 4:30pm (CST), or email productremovalinfo@nurseassist.com
Initiated
2023-11-06
Root cause
Process control
Distribution
US Nationwide Distribution: SC AZ, CA, FL, GA, HI, IA, IN, KS, KY, MA, ME, MN, MS, NC, ND, NE, NJ, NV, NY, OH, PA, RI, TX, WA, WI, AK, AL, AR, CO, CT, DC, ID, IL, LA, MD, MI, MO, MT, NH, NM, OK, OR, PR, SD, TN, UT, VA , WI, WV, WY. OUS (foreign) distribution to: Canada
Recall number
Z-0513-2024
Recall status
Open, Classified
CDRH record ID
204367
FDA classified
2023-12-20
Recall event number
93434

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗