Skip to content

Public FDA record

Z-0514-2020: Stryker X-Change Medial Wall Mesh Small (6-Petal)-Intended to be implanted to replace a hip joint REF 094

Recall record Z-0514-2020 for Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0514-2020
Company in FDA record
Howmedica Osteonics Corp.
FDA record creation date
Nov 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Boxes of the 6-Petal X-Change Wall Mesh Small (Part # 0942-8-025) contain the 4-Petal X-Change Wall Mesh Small (Part # 0942-8-015)

Source facts

Action
Stryker issued An Urgent Field Safety Notice (FSN), PFA 2197654, was issued to all affected OUS consignees at country level on October 03, 2019. Action to take: Actions to be taken by the Customer/ User: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be aware within your organization. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. A response is required, even though you may not have any physical inventory on site anymore. 3. Quarantine and discontinue use of the recalled devices. 4. Complete customer response form and return the form and any affected devices to the following address. (Please complete this form regardless of whether your organization has product to return. This will preclude the need for Stryker to send any reminder notice.) ATTN: David Nihill, Limerick Finished Goods Centre C/O Flextronics Global Services Ireland Unit 1 Raheen Business Park Raheen, Limerick.
Initiated
2019-10-03
Root cause
Employee error
Terminated
2022-04-21
Distribution
Foreign: Australia, France, Finland, Germany, Italy, Japan, New Zealand, Netherlands, Norway, Romania, Spain, Sweden and UK
Recall number
Z-0514-2020
Recall status
Terminated
CDRH record ID
177408
FDA record created
2019-11-22
Recall event number
84148

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗