Z-0514-2021: Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supp

Recall record Z-0514-2021 for Invacare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0514-2021
Company in FDA record
Invacare Corporation
FDA record creation date
Dec 1, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.

Source facts

Action
Invacare issued Urgent Notification letter on 10/30/20 to providers and consumers via email stating reason for recall, health risk and action to take: Visit www.Invacare.com/RC928 within five (5) business days of receiving this notice. Enter the requested identification information to access the list of affected device serial numbers sold to you and instructions to conduct this field correction. You will be prompted to provide responses to information requested regarding this field correction and affected units. 2. Review your existing stock to locate the affected devices. 3. Review the Repair Instructions provided on the www.Invacare.com/RC928 website. 4. Order field correction service kits and conduct field correction repair activities on affected devices during their next regularly scheduled maintenance service. Details on how to order the service kits are provided on the website www.Invacare.com/RC928. 5. Field correction repairs must be completed on affected units before October 30, 2021. If you currently use a service provider for maintenance and repair, they can execute this field action service. 6. You are responsible for ensuring that this field correction is conducted with your customers. Customers who own affected units should be contacted and informed about this field correction. We recommend that you provide them with the Consumer Letter provided on the www.Invacare.com/RC928 website. Invacare provided copies of the US and Canada notification letters to Invacare Europe for their reference and distribution.
Initiated
2020-10-30
Root cause
Device Design
Terminated
2024-12-31
Distribution
Worldwide distribution. US Nationwide, Canada, Australia, China, Germany, Guam, New Zealand, and Northern Mariana Islands
Recall number
Z-0514-2021
Recall status
Terminated
CDRH record ID
184169
FDA record created
2020-12-01
Recall event number
86604

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗