Z-0514-2022: SonarMed AirWave Monitor Kit (Monitor, Sensor Cable, Power Cord), Item Code AW-M0001. used with endotrac

Recall record Z-0514-2022 for Covidien, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0514-2022
Company in FDA record
Covidien, LP
FDA record creation date
Jan 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable disconnect and reconnect may cause the monitor to discontinue monitoring and transition into the calibration screen without an audible alarm.

Source facts

Action
Urgent Medical Device Recall were distributed to customers beginning 12/22/21. Actions you should take: " Immediately discontinue use of the affected SonarMed AirWave" monitors. Remove the affected devices listed on Attachment A from clinical service. " Notify all personnel in all care environments in which the affected SonarMed AirWave" monitors are used about this medical device recall. " If your facility has distributed affected SonarMed AirWave" monitors to other persons or facilities, please promptly forward a copy of this letter to those recipients. " Complete the attached Recalled Product Return Form even if you do not have inventory. " Customers are requested to return the item codes listed on Attachment A as described in this notice. Actions being taken by Medtronic: " Medtronic is implementing manufacturing process improvements to address the manufacturing assembly issue and is developing a software update to address the identified software anomaly. This software is projected to be available in Spring 2022 and will be factory installed on all newly manufactured SonarMed AirWave" monitors. Purchased directly from Medtronic: Customer with inventory: 1. Complete the Recalled Product Return Form and fax or email to Medtronic. 2. Ship affected product with Retuned Goods Authorization (RGA) number provided by Customer Service to: Medtronic Attn: Field Returns 195 McDermott Road North Haven, CT 06473. Customer with zero inventory: Complete form and check the box indicating no inventory and fax or email to Medtronic. Where to send the completed form: Fax to 800-895-6140 or email to rs.covidienfeedbackcustomerservice@medtronic.com Credit for returned affected product will be issued based on the RGA number. Exception: Customers with zero inventory, fax to 651-367-2624 or email to rs.gmbfcamitg@medtronic.com Purchased from a distributor: Customer with inventory: 1. Complete the Recalled Product Return Form and fax or email to Medtron
Initiated
2021-12-22
Root cause
Process control
Distribution
The devices were distributed to the following US states: CA, CO, FL, IL, KY, MA, OH, TN, and TX.
Recall number
Z-0514-2022
Recall status
Open, Classified
CDRH record ID
191020
FDA record created
2022-01-19
Recall event number
89283

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗