Z-0516-06: Clinical Chemistry Multiconstituent Calibrator--- List number: lE65-02

Recall record Z-0516-06 for Abbott Laboratories Diagnostic Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0516-06
Company in FDA record
Abbott Laboratories Diagnostic Div
FDA classification date
Feb 14, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The two Multiconstituent Calibrator (MCC) lot numbers 19906M200 and 20879M100 failed to meet product performance expectations throughout the labeled expiration period, as a result of the shortened expiration dating, two lots are now expired. The expiration date for Lot number 26674M100 will be revised from 2006-07-31 to 2006-01-31.

Source facts

Action
A Device Correction letter identifying the changes being made to the expiration dating for lot numbers 19906M200, 20879MI00 and 26674M1O0 along with a Customer Reply form will be distributed via FedEx to all AEROSET and ARCHITECT c8000 customers which utilize the Clincial Chemistry Multiconstituent Calibrator on December 27,2005. Outside the U.S. each country organization will need to determine the appropriate method for the effectiveness check.
Initiated
2005-12-23
Root cause
Other
Terminated
2012-04-24
Distribution
Nationwide plus worldwide in Mexico,Venezuela, Chile, Brazil, San Salvador, Singapore, Korea, Trinidad, Grand Cayman, Guatemala, Columbia, Uruguay, Puerto Rico, Canada, Hong Kong, Australia, Bermuda, Jamaica, Costa Rica, Ecuador, Argentina, Dominican ReD., Germanv, Thailand, New Zealand, Bahamas, Panama, & Paraguay.
Recall number
Z-0516-06
Recall status
Terminated
CDRH record ID
43861
FDA classified
2006-02-14
Recall event number
34435

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗