Skip to content

Public FDA record

Z-0516-2013: Motorola Solutions PCS3000/CS3070 Laser Utility/Peripheral Laser Products

Recall record Z-0516-2013 for Motorola Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0516-2013
Company in FDA record
Motorola Inc
FDA classification date
Jan 30, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered the CS3000/CS3070 was manufactured using the laser power settings for a Class 2 laser device instead of the Class 1 power levels as specified on the product label.

Source facts

Action
Motorola Solutions, Inc. agrees to provide instructions for downloading and installing a software update from the internet, which will reduce the laser power to IEC Class 1 AELs. Purchasers will be provided an internet site and instructions for notifying Motorola Solutions, Inc. that the software installation and power reduction were successful, thereby completing the corrective action. corrections to all units is anticipated to be completed by May 2013. I. Notifications to all dealers and purchasers must be made, and the FDA NY District Office Recall Coordinator is to be included in the notifications. 2.Correction of all units to be completed by May 2013. 3.All products will be brought into compliance with the laser performance standard at no cost to the purchaser. For questions regarding this recall call 631-738-3474.
Initiated
2012-04-27
Root cause
Radiation Control for Health and Safety Act
Terminated
2018-06-29
Distribution
Nationwide distribution
Recall number
Z-0516-2013
Recall status
Terminated
CDRH record ID
109265
FDA classified
2013-01-30
Recall event number
61870

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗