Z-0516-2020: UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage

Recall record Z-0516-2020 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0516-2020
Company in FDA record
Beckman Coulter, Inc.
FDA record creation date
Nov 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.

Source facts

Action
Consignees were notified via email and mail with an Urgent Medical Device Recall letter, dated 10/02/2019. Four consignees were notified with a letter dated 10/18/2019. The letter identified affected product, stated the issue, provided action to be taken to determine if consignee is affected, and stated that a hardware change was in progress to eliminate the issue. A response form was asked to be returned. Questions can be directed to Beckman Coulter Customer Support Center at 800-526-7694.
Initiated
2019-10-07
Root cause
Under Investigation by firm
Terminated
2021-04-27
Distribution
Worldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-0516-2020
Recall status
Terminated
CDRH record ID
177542
FDA record created
2019-11-22
Recall event number
84186

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗