Z-0516-2025: RENASYS EDGE The RENASYS EDGE pump is indicated for patients who would benefit from a suction pump (NPWT

Recall record Z-0516-2025 for Smith & Nephew Medical, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0516-2025
Company in FDA record
Smith & Nephew Medical, Ltd.
FDA classification date
Nov 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to issues with battery depletion and inability to recharge if the batteries are not appropriately maintained.

Source facts

Action
On 10/02/2024, the firm sent via overnight mail an "Urgent Medical Device Correction Notice" to customers informing them of an issue associated with the RENASYS EDGE batteries if not appropriately maintained and charged, the batteries may become depleted and be unable to recharge. Customers are instructed to: 1. Please inspect your inventory and locate any RENASYS. EDGE devices from the listed product and serial numbers enclosed with this Urgent Medical Device Correction Notice, and quarantine them immediately. a. If you are a sales rep, please work with sales leadership to have device upgraded or returned. b. If you are a distributor/DME, your Care Continuum Director will contact you with next steps once you submit the completed Acknowledgment Form. Note: Based on the regulatory classification and degree of hazard, this field action extends to the retail level. Therefore, it is not required to communicate this notice to the patient at the consumer level. 2. Complete Correction Acknowledgement Form and submit to fieldactions@smith-nephew.com. Firm's contact information: Smith & Nephew, Inc. Global Field Actions 1450 E Brooks Road Memphis, TN 38116 www.smith-nephew.com T: + 1 901 396 2121 T: 1 800 821 5700 (USA toll free)
Initiated
2024-10-02
Root cause
Process control
Distribution
U.S.: AL, CA, CO, FL, HI, IN, KY, NE, NY, PA, SC, TN and WI O.U.S.: Canada
Recall number
Z-0516-2025
Recall status
Open, Classified
CDRH record ID
210584
FDA classified
2024-11-22
Recall event number
95558

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗