Open original FDA record ↗Recall record
- FDA identifier
- Z-0516-2025
- Company in FDA record
- Smith & Nephew Medical, Ltd.
- FDA classification date
- Nov 22, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to issues with battery depletion and inability to recharge if the batteries are not appropriately maintained.
Source facts
- Action
- On 10/02/2024, the firm sent via overnight mail an "Urgent Medical Device Correction Notice" to customers informing them of an issue associated with the RENASYS EDGE batteries if not appropriately maintained and charged, the batteries may become depleted and be unable to recharge. Customers are instructed to: 1. Please inspect your inventory and locate any RENASYS. EDGE devices from the listed product and serial numbers enclosed with this Urgent Medical Device Correction Notice, and quarantine them immediately. a. If you are a sales rep, please work with sales leadership to have device upgraded or returned. b. If you are a distributor/DME, your Care Continuum Director will contact you with next steps once you submit the completed Acknowledgment Form. Note: Based on the regulatory classification and degree of hazard, this field action extends to the retail level. Therefore, it is not required to communicate this notice to the patient at the consumer level. 2. Complete Correction Acknowledgement Form and submit to fieldactions@smith-nephew.com. Firm's contact information: Smith & Nephew, Inc. Global Field Actions 1450 E Brooks Road Memphis, TN 38116 www.smith-nephew.com T: + 1 901 396 2121 T: 1 800 821 5700 (USA toll free)
- Initiated
- 2024-10-02
- Root cause
- Process control
- Distribution
- U.S.: AL, CA, CO, FL, HI, IN, KY, NE, NY, PA, SC, TN and WI O.U.S.: Canada
- Recall number
- Z-0516-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210584
- FDA classified
- 2024-11-22
- Recall event number
- 95558