The transducer used to compensate for pressure changes was mounted backwards on the printed circuit board assembly. The module will not compensate for pressure changes and CO2 readings will be inaccurate at higher altitudes.
Source facts
Action
On 7/20/05 Spacelabs called the consignees that received the units to notify them of the hazard and also the schedule a service call to repair the affected capnography modules.
The firm will confirm that all consignees have been updated with the appropriate PCBA and will audit the device recall records for completion.
Initiated
2005-07-20
Root cause
Other
Terminated
2006-02-17
Distribution
8 units shipped to France, 1 unit to Brussels, Belgium, and
2 units to a medical center in WY.
Recall number
Z-0517-06
Recall status
Terminated
CDRH record ID
40875
FDA classified
2006-02-16
Recall event number
32815