Z-0517-2021: MiniMed 630G with SmartGuard, insulin pump, Ref/Model # MMT-1754K, MMT-1714K, MMT-1755K - Product Usage:

Recall record Z-0517-2021 for Medtronic Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0517-2021
Company in FDA record
Medtronic Inc.
FDA record creation date
Dec 2, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Missing solder battery connection, which could interrupt power of insulin pump.

Source facts

Action
On April 30, 2020, the firm mailed a "Urgent Medical Device Recall" letter to patients. The firm also sent a letter to Health Care Professionals, (HCP) "Urgent Medical Device Recall" that they read this notice and confirm receipt that the customer read and understood the instructions provided in this letter. The letter notified customers that the insulin pumps had an issue on the battery connection that did not meet firm's quality specifications. If power is interrupted on a patients insulin pump, then insulin is suspended, which may lead to hyperglycemia (high blood sugar). There have been no reported complaints from customers as a result of this issue. The firm identified the pump models, serial numbers of the affected devices. In addition to informing patients about the recall, the letter asked patients to do the following: (1) Record current pump settings. Follow the System User Guide to set up new pump that is enclosed in the box. (2) After successfully programming the replacement pump, start using replacement pump. (3) Return affected pump using the included shipping materials. Contact 24-Hour Technical Support at 1-877-585-0166, the instructions provided in this letter. The firm sent a letter to Health Care Professionals, (HCP) "Urgent Medical Device Recall" that they read this notice and confirm receipt that the customer read and understood lead to hyperglycemia (high blood sugar). There have been no reported complaints from customers as a result of this issue. The firm identified the pump models, serial numbers of the affected devices.
Initiated
2020-04-30
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - U.S. Nationwide Distribution in the state of MD and the countries of Canada, Netherlands, Norway. ***Model 630G distributed in Canada and Model 670G distributed in US ***MODEL640G Distributed OUS ONLY*** OUS: Australia, Canada, Netherlands and Norway.
Recall number
Z-0517-2021
Recall status
Open, Classified
CDRH record ID
181607
FDA record created
2020-12-02
Recall event number
85711

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗