Open original FDA record ↗Recall record
- FDA identifier
- Z-0517-2024
- Company in FDA record
- Bebig Isotopentechnik Gmbh
- FDA classification date
- Dec 8, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to software malfunction, numerical values may be rounded resulting in inaccurate measurements.
Source facts
- Action
- On October 17, 2023, Eckert & Ziegler BEBIG GmbH issued a "Urgent Field Safety Notice" via email. Eckert & Ziegler asked consignees to take the following actions: 1. If the Length Unit is set to millimeters, user must carefully review and, if necessary, adjust the Image Sequence Plane Separation value during import when working with slice thicknesses that have two decimals, as shown in Figure 1. Users must consider that this approach is not intrinsically safe, therefore solution 1 or 2 is highly recommended 2. If you have used mm as length unit and may have used two decimal numbers in mm for slice thickness, carefully review the patient cases and act accordingly. 3. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the devices have been transferred to. Please reply to this email by 20th of October 2023 confirming that you have received and understood this information and that you have forwarded it to the people which already have received the named products or will receive them.
- Initiated
- 2023-10-17
- Root cause
- Software Design Change
- Distribution
- US: MD OUS: Germany
- Recall number
- Z-0517-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204260
- FDA classified
- 2023-12-08
- Recall event number
- 93399