Z-0518-2020: Site~Rite 8 Ultrasound Systems-intended for diagnostic ultrasound imaging of the human body. Site~Rite 8

Recall record Z-0518-2020 for Becton Dickinson & Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0518-2020
Company in FDA record
Becton Dickinson & Company
FDA record creation date
Nov 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The indicator may show available charge when the battery is close to depletion. As a result, the Site~Rite 8 has the potential to abruptly shut down, without advance warning to the user, when running on battery power only

Source facts

Action
BD issued to US consignees via FedEx beginning November 4, 2019 Urgent Medical Device Notification, stating reason for recall , health risk and action to take: Until the issue is resolved, it is recommended that all Site~Rite 8 Ultrasound Systems be operated using AC power at all times during use in a clinical procedure. If use of AC power is not practical for your clinical setting, we recommend the use of an external re-chargeable battery. The listing of batteries have been qualified for use with the Site~Rite 8 Ultrasound System and are available for purchase through common retailers were identified. If you have experienced an abrupt shutdown, please contact 1-800-296-4146 Opt. 1 to coordinate with BD to replace your battery or provide you with an external battery. 2. Please share this medical device notification with all users of the Site~Rite 8 Ultrasound System at your facility to ensure they are also aware of the issue. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form so that BD may acknowledge your receipt of this notification. 4. If you choose to purchase one of the above batteries independently, please contact BD Customer Service at capital_support@bardaccess.supportsystem.com. Please include a copy of the purchase receipt, BD account number and serial number of affected unit to arrange for reimbursement. 5. If the above batteries are not available, please contact BD Customer Service for additional battery options. Contact Information: 1-800-290-1689 Monday Friday between 8:00am and 5:00pm (EST) in the United States.
Initiated
2019-11-04
Root cause
Under Investigation by firm
Terminated
2020-11-19
Distribution
Nationwide Foreign: UK, Belgium, Australia, Japan, Canada, Taiwan, Colombia, Chile, Philippines, Thailand, Brazil, Hong Kong, Chile, Argentina, Mexico, India, Singapore
Recall number
Z-0518-2020
Recall status
Terminated
CDRH record ID
177853
FDA record created
2019-11-23
Recall event number
84262

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗