Open original FDA record ↗Recall record
- FDA identifier
- Z-0518-2024
- Company in FDA record
- Riverpoint Medical, LLC
- FDA classification date
- Dec 8, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.
Source facts
- Action
- On 10/25/2023, the firm emailed its customer an "URGENT: MEDICAL DEVICE CORRECTION" Letter informing them that the Iconix Knotless Anchor 's Instruction for Use is insufficient to prevent/avoid inserter buckling and insertion failure. The customers were instructed to do the following: - Immediately examine your inventory. If you have the part numbers listed above, notify end user with additional instructions for use. Users may continue to use field inventory once the acknowledgement letter has been completed. - Complete and return the response form as soon as possible to quality@rpmed.com. Mark the return shipments with RGA#2081- PDX. - Provide field correction information to any customers to whom you may have distributed or forwarded affected products. You may include a copy of the recall notice with your customer notification. - A copy of the notification letter and of IFU may be posted on or near the affected products, so users are aware of the additional instructions.
- Initiated
- 2023-10-25
- Root cause
- Component design/selection
- Distribution
- U.S. Nationwide distribution in the state of CA.
- Recall number
- Z-0518-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204240
- FDA classified
- 2023-12-08
- Recall event number
- 93390