Open original FDA record ↗Recall record
- FDA identifier
- Z-0520-2008
- Company in FDA record
- Vital Images, Inc.
- FDA classification date
- Jan 31, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Standard Uptake Values (SUV) display/update problems; in the way Standard Uptake Values (SUV) are displayed in 3D/MPR views. The problem arises when a user draws a selected region of interest in order to calculate the SUV statistics of the region. If the user then changes the slice shown by scrolling through additional slices, the original SUV measurements will remain on the screen and do not up
Source facts
- Action
- Customers were notified via telephone (11/7 -11/8/2007) and issued an Urgent Medical Device recall notification letter on 11/8/07. The letter described the problem and product involved. The letter also informed customers they they will provide a free software patch within the next 6-8 weeks.
- Initiated
- 2007-11-08
- Root cause
- Process control
- Terminated
- 2011-12-17
- Distribution
- Worldwide-USA including states of SC, MO, MI, WI, NY, FL, CO, CA, PA, and WA, and countries of England and The Netherlands
- Recall number
- Z-0520-2008
- Recall status
- Terminated
- CDRH record ID
- 66214
- FDA classified
- 2008-01-31
- Recall event number
- 45820