- FDA identifier
- Z-0520-2012
- Company in FDA record
- Del Mar Reynolds Medical, Ltd.
- FDA classification date
- Jan 11, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
When Lifescreen is left open and a new recording is loaded from Sentinel Cardiology Information Management System, customers can get corruption in the data or a wrong ECG after the first 32 hours. Both recordings have to be longer than 32 hours for this to happen.
Source facts
- Action
- Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated November 11, 2011 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. Customers are advised to close Lifescreen every time before loading another recording from the Sentinel database. Spacelabs is in the process of upgrading software and will contact customers to arrange for an update to affected systems. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
- Initiated
- 2011-11-11
- Root cause
- Software design
- Terminated
- 2012-06-05
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AR, CA, CO, CT, FL, GA, KY, LA, MA, MD, MI, MN, NC, NY, OH, TN, TX, UT, WV, and PUERTO RICO and the countries of AUSTRIA, BELGIUM, CHILE, CHINA, COLOMBIA, COSTA RICA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, ISRAEL, ITALY, KUWAIT, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, Slovakia, SPAIN, SWEDEN, SWITZERLAND, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM
- Recall number
- Z-0520-2012
- Recall status
- Terminated
- CDRH record ID
- 105891
- FDA classified
- 2012-01-11
- Recall event number
- 60592